| Pure Steam Sterilizer |
|
|
|
| |
|
|
|
|
Specially designed to produce sterile & pyrogen free steam at 3 kg/cm² (142°c) |
|
|
|
|
|
Used for steam sterilization of big vessels, pipeline, autoclaves etc. |
|
|
|
|
|
Provided with automatic control panel board with fault detection. |
|
|
|
|
|
Auto function with your autoclave. |
|
|
|
|
|
All contact parts in S.S 316 |
|
|
|
|
|
Sterile steam passes through lal test. |
|
|
|
|
|
|
| Scope of Supply |
|
|
|
|
|
|
Feed water pump in s.s.316 'grunfos' make. |
|
|
|
|
|
Conductivity meter for fee water. |
|
|
|
|
|
Automatic panel board. |
|
|
|
|
|
Control valve for throttling the steam. |
|
|
|
|
| Description |
|
|
|
|
|
|
|
Pure steam generator is designed to produce Pyrogen Free Sterile for Steam Sterilization for Pharmaceutical and Biotech industries. This is more reliable than Boiler Clean Steam. The Pure Steam Generator steam meets the IP/BP and standard. |
|
|
|
|
|
| Operation |
|
|
|
|
|
|
|
|
D.M.Water is Pre-Heated in Pre-Heaters (2 or 3 nos.) before going to main column where it gets converted into pyrogen free sterile steam by taking heat from boiler steam at 6 kg/cm2 pressure. It is provided with S.S.control valve for throttling the steam. |
|
|
|
|
|
| Operating Parameters |
|
|
|
|
|
|
|
|
|
Boiler Steam : Minimum 6 kg/cm2 pressure at 165 C. |
|
|
|
|
|
|
|
|
|
Feed Water : D.M. Water with conductivity less then 5 Micromhos. |
|
|
|
|
|
|
|
|
|
Pure Steam : 3 kg/cm2 at 142 C as per IP/BP Standard. |
|
|
|
|
|
|
|
|
|
Lal test : Passes through lal test. |
|
|
|
|
|
|
|
|
|
Power : 5 kw for 100, 200 kgs. & 7 kw for 300 & 500 kgs. |
|
|
|
| Application |
|
|
|
|
|
|
|
|
|
|
|
|
|
It is used for Sterilization and Sanitization of Pharma Equipments, Autoclaves, Piping & Tanks. |
| |
|
|
|
|