Bio-Medical 2B2 Model  
Reliance Instruments Corporation The vacuum autoclave / Sterilizer is designed for strilization of regulated bio- Medical Waste generated from the hospitals for destruction of bacteria and related organism. The equipment produced for bio-medical waste management operate in accordance with the specifications laid down by central pollution control board. The sterilizer is required to handle bio-medical waste of 10-100 kg/per cycle of sterilization is able to perform approx 5 cycles per shift of 8 hours in a day.
 
         
No of Bed Waste Generation(opproximate) Model No
Up to 50 10 kg/hr BMW 10
50-200 25 kg/hr BMW 25
300-500 50 kg/hr BMW 50
500-800 75 kg/hr BMW 75
800 and above 100 kg/hr BMW 100
         
The equipment to treat medical waste shall be operated in accordance with the following minimum requirement specified by the pollution control board.  
         
Construction and Features  
         
The equipment is designed with low chamber floor height for easy loading and unloading.  
       
0.2 micron special filter is used for air to equivalise chamber pressure after drying.  
       
The chamber will have sturdy gasket capable to withstand double the operating pressure  
       
Control and monitoring system  
         
Microprocessor computer system and printer shall be provided for controlling and monitoring the sterilization cycle with real time printer with 18 column giving cycle progression each minute indicating stage of cycle ,Data ,Day, Load Identification number, temperature and pressure or vacuum ,errors and failures, unchangeable date and time record in during each cycle to avoid any tempering by operator
 
       
Fully automatic sliding door pneumatically operated with fully advanced programmable logic controller with 80 column full status recording printer.
 
       
Rapid Cooling Systems for liquids in bottles and PVC pouches.  
       
N .B the condensate left out with other residues will be fully treated for odorless.  
       
Operating parameters for Waste Management(Autoclaves)  
         
1.Operating temperature 135° C
2.Operating pressure 2.2 kg/cm2 or 33psi
3.Residance time for above two parameters 130 minutes
4.Drying through Vacuum 20 min. after (sterilization)
5. No of pre- vacuum pulses 1 to 5
         
Specification  
         
Chamber SS 316
Jacket SS 304
Insulation 75 mm
Insulation Cover Powder coated
Working Pressure 32Psi
Max Working Temperature !37°C
Mode of Heating Electrical/ External Piped Steam
Doors Locking System Radial / Sliding / Cam locking and Tongue and Groove System
Shape Inside Cylendrical/ Rectangular
Water Filling Manual / Intelligence auto Filling
Mode of operation Manual mutiport /Valves Optional Micro Controler/PLC /SCADA
Safety Device Automatic Door Locking System, Safety valve, Water low Cut off Device Etc….
Accessories Transfer trolleys, Temperature Recorders, Validation Ports(optional)
Filters 0.5 Micron during Drying.
Automatic sequence Gravity, Express, Leak Rate Test and Bowie-Dick Test With FO Value Calculation
High Vacuum model Upto 760mm/Hg with power full vacuum pump 2-4 Pulsing systems
Steam temperature 145 °C Before entering Chamber
Pre heater For exhaust out let at 150°C
Data logging Through Serial port printer or computer
Tilting Chamber pneumatic Tilting Optional.
         